PS
Pressure transmitters with piezoresistive sensing principle
Service intended: Relative and absolute pressure measurement of low and medium pressure ranges, gaseous and liquid media, the flush diagram type also for highly viscous and crystallizing pressure media.

hr
 |
|
Downloads |
Technical Details
Operation Manual
Operation Manual EX
Operation Manual IP68
Operation Manual 0-100mV |
|
Technical Data |
|
Case |
|
Stainless steel 304S |
|
Degree of protection |
|
IP 65 per EN 60529 / IEC 529 |
|
Pressure connection |
Stainless steel 316 Ti
G1/2B per EN 837-1
G3/4B with flush diaphragm
G1/4 female |
|
Sensing element |
|
st. st. 316L, Measuring bridge: piezoresistive |
|
Pressure and overpressure ranges |
|
see Technical Details page 05.0102 |
|
Electrical connection |
|
L-plug EN175301-803 Form A |
Signal output Power supply |
|
4 ... 20 mA, 2-wire Ub = 20 ... 30 V DC |
|
Signal output Power supply |
4 ... 20 mA, 2-wire Ub = 12 ... 30 V DC
0 ... 20 mA, 3-wire |
|
Load |
Rb [Ohm] : (Ub [ V ]-12 [ V ] )/0,025 [ A ]
Rb [Ohm] : (Ub [ V ]-20 [ V ] )/0,025 [ A ]  |
|
Response time |
|
t < 10 ms |
|
Power consumption |
|
P max = 0,75 W |
Bleeder resistor  |
|
RAb >= 1 MOhm |
|
Permissible temperatures Tmin / Tmax |
Storage temperature - 40 ... +100°C
Ambient temperature - 20 ... + 70°C
II 2G Ex ia IIC T4 - 20 ... + 70°C
II 2G Ex ia IIC T6 - 20 ... + 40°C
Medium temperature - 25 ... + 80°C
Compensated range - 10 ... + 80°C |
|
Accuracy |
|
Linearity + hysteresis + repeatability < 0,3 % FS |
|
Temperature error |
Mean TK of zero < 0,2 % FS / 10 K
Mean TK of range < 0,2 % FS / 10 K |
|
Weight |
|
0,27 kg |
|
Optional extras |
|
see Technical Details page 05.0305 |
|
Digital indicator |
|